Sterilization (microbiology) Study Guide
Study Guide
📖 Core Concepts
Sterilization: Complete elimination (or irreversible inactivation) of all microorganisms, spores, prions, and viruses. Objects become sterile or aseptic.
Sterility Assurance Level (SAL): Probability that a unit remains non‑sterile after processing (e.g., SAL = 10⁻⁶ means 1 failure in 1,000,000 units).
D‑value: Time (or dose) needed to achieve a 1‑log (90 %) reduction in viable count; \(D = 1/k\).
First‑order kill kinetics: Microbial death rate proportional to the surviving population, \( \frac{dN}{dt} = -kN \).
Overkill method: Sterilize for a margin beyond the measured bioburden to guarantee the target SAL.
Aseptic technique: Practices that keep sterile items free from contamination (sterile garments, gloves, cleanroom environment).
📌 Must Remember
SAL requirement for high‑risk medical products: ≤ 10⁻⁶ (FDA).
Steam autoclave parameters: 121 °C – 134 °C for 3–30 min (gravity‑displacement) or 18 min at 134 °C for prions.
Dry heat: 160 °C ≥ 2 h (standard) or 190 °C for 6 min (unwrapped).
Ethylene oxide (EO) gas: 30 °C – 60 °C, RH > 30 %, 200–800 mg/L, several hours.
Membrane pore sizes: 0.22 µm → most bacteria; 20–50 nm → viruses.
Biological indicator for steam: Geobacillus stearothermophilus spores.
Radiation dose for sterilization: ≥ 10 MeV to induce radioactivity (not used); typical sterilization energies are below this, so items stay non‑radioactive.
🔄 Key Processes
Heat‑based microbial kill (first‑order)
\(N = N0 e^{-kt}\) → \(N = N0 10^{-t/D}\).
Determine D‑value at a reference temperature, then adjust for other temps using Arrhenius: \(k = A e^{-Ea/(RT)}\).
Steam sterilization cycle
Load → pre‑vacuum (or gravity‑displacement) → heat to 121–134 °C → hold for prescribed time → depressurize → dry.
EO gas sterilization
Load → set temperature, RH, and concentration → expose for required duration → aeration to remove residual EO.
Sterile filtration
Pre‑filter → select appropriate membrane (0.22 µm or 20–50 nm) → pass fluid tangentially → monitor pressure/flow → perform integrity test.
🔍 Key Comparisons
Steam vs. Dry Heat
Moist heat denatures proteins at lower temps (121 °C) → faster; dry heat requires higher temps (≥ 160 °C) and longer exposure.
EO Gas vs. Hydrogen Peroxide
EO penetrates porous plastics, low temp, longer cycle, toxic residues.
H₂O₂ (vapour) faster, no toxic residues, but limited penetration and material incompatibility (cellulose, nylon).
UV vs. Ionizing Radiation
UV: surface‑only, non‑penetrating, low dose, ineffective in shadows.
Ionizing (γ, e‑beam, X‑ray): deep penetration, higher dose, can sterilize bulk loads.
⚠️ Common Misunderstandings
“Sterilization = Disinfection” – Sterilization eliminates all life; disinfection only reduces microbial load.
“All prions are killed by standard autoclave cycles” – Prions need longer times (≥ 60 min at 121 °C) or higher temps (134 °C ≥ 18 min).
“Filtration removes all viruses” – Only nanofiltration (≤ 50 nm) can reliably remove viruses; 0.22 µm filters do not.
🧠 Mental Models / Intuition
Log‑kill “stairs”: Each D‑value is one step down a 10‑fold ladder; 3 D‑values = 99.9 % kill, 6 D‑values = 99.9999 % kill (≈ SAL 10⁻⁶).
Temperature‑time trade‑off (FT‑T): Higher temperature → smaller D‑value → fewer minutes needed (inverse exponential via Arrhenius).
Barrier vs. Inactivation: Heat/chemicals inactivate microbes; filtration physically blocks them.
🚩 Exceptions & Edge Cases
Prion resistance: Require extended high‑temperature steam cycles; many chemical agents (including aldehydes) are ineffective.
Material incompatibility: Dry heat can cause oxidation; EO may degrade certain polymers; H₂O₂ can embrittle nylon.
Air‑dry cycles: Gravity‑displacement autoclaves may leave air pockets in complex loads → risk of incomplete sterilization.
📍 When to Use Which
Heat‑stable metal tools → Steam sterilization (fast, reliable).
Heat‑sensitive plastics, electronics → EO gas or VHP (choose EO for deep penetration, VHP for quicker cycle & no toxic residues).
Liquid pharmaceuticals that cannot be heated → Sterile filtration (0.22 µm for bacteria, nanofiltration for viruses).
Large bulk items or pallets → Gamma or high‑energy X‑ray (deep penetration).
Surface decontamination of workspaces → UV germicidal lamps (short‑range, no residues).
👀 Patterns to Recognize
“D‑value decreases as temperature rises” → look for Arrhenius‑type statements.
“Biological indicator spore = most resistant organism” → expect G. stearothermophilus for steam, similar for other methods.
“Cycle time + temperature = target SAL” → if the question gives time and temp, compute log‑kill steps to verify SAL.
“Pore size vs. target organism” → bacteria → 0.22 µm; viruses → ≤ 50 nm.
🗂️ Exam Traps
Distractor: “Steam sterilization kills all prions at 121 °C for 15 min.” – Wrong; prions need longer or higher temperature.
Distractor: “EO sterilization works at any humidity.” – Incorrect; RH must be > 30 % for effective EO diffusion.
Distractor: “UV can sterilize opaque liquids.” – UV cannot penetrate opaque or turbid media.
Distractor: “A 0.22 µm filter guarantees viral removal.” – Only nanofiltration (20–50 nm) removes viruses.
Distractor: “Higher SAL (e.g., 10⁻⁴) is safer than 10⁻⁶.” – Lower numeric SAL indicates higher assurance (10⁻⁶ is safer).
---
Keep this guide handy; each bullet is an exam‑ready fact or decision rule.
or
Or, immediately create your own study flashcards:
Upload a PDF.
Master Study Materials.
Master Study Materials.
Start learning in seconds
Drop your PDFs here or
or